Skip Ribbon Commands
Skip to main content
Sign In

26482 CRRT circuit IC

CRRT circuit IC
Test Code: IC-CIR
Test Components

​Ionized Calcium

Specimen Requirements
Specimen TypePreferred Container/TubeAcceptable Container/TubeSpecimen VolumeSpecimen Minimum Volume
(allows for 1 repeat)
​Whole Blood​Low or balanced heparin blood gas syringe​Full Li Heparin no gel tube​Full syringe or tube​Full syringe or tube
Collection Processing Instructions

Marshfield ONLY:  Collect heparinized whole blood anaerobically in a low- or balanced-heparin blood gas syringe or LI Heparin no gel tube.  Upon receipt in the lab, do not centrifuge.  Deliver to the blood gas testing section for analysis.

Specimen Stability Information
Specimen TypeTemperatureTime
​Whole Blood- syring/non-gel GTT​Refrigerated​4 hours
Ambient​30 minutes​
Rejection Criteria

​Specimens exposed to air

​Plasma Separator tubes
​Gross hemolysis

 

Performing Laboratory Information
Performing LocationDay(s) Test PerformedAnalytical TimeMethodology/Instrumentation
​Marshfield​Sunday through Saturday​Less than 1 hur​IL GEM 4000
Reference Range Information

No reference range established for this analyte/sample type

Test Components

​Ionized Calcium

Ordering Applications
Ordering ApplicationDescription
​Cerner​IC-CIR
If the ordering application you are looking for is not listed, contact your local laboratory for assistance.
Specimen Requirements
Specimen TypePreferred Container/TubeAcceptable Container/TubeSpecimen VolumeSpecimen Minimum Volume
(allows for 1 repeat)
​Whole Blood​Low or balanced heparin blood gas syringe​Full Li Heparin no gel tube​Full syringe or tube​Full syringe or tube
Collection Processing

Marshfield ONLY:  Collect heparinized whole blood anaerobically in a low- or balanced-heparin blood gas syringe or LI Heparin no gel tube.  Upon receipt in the lab, do not centrifuge.  Deliver to the blood gas testing section for analysis.

Specimen Stability Information
Specimen TypeTemperatureTime
​Whole Blood- syring/non-gel GTT​Refrigerated​4 hours
Ambient​30 minutes​
Rejection Criteria

​Specimens exposed to air

​Plasma Separator tubes
​Gross hemolysis

 

Test Components

​Ionized Calcium

Reference Range Information

No reference range established for this analyte/sample type

For more information visit:
Performing Laboratory Information
Performing LocationDay(s) Test PerformedAnalytical TimeMethodology/Instrumentation
​Marshfield​Sunday through Saturday​Less than 1 hur​IL GEM 4000
For billing questions, see Contacts
For most current information refer to the Marshfield Laboratory online reference manual.