Assay results should only be used in conjunction with information available from the clinical evaluation of the patient and other diagnostic procedures.
In rare cases, some individuals can develop antibodies to mouse or other animal antibodies (often referred to as human anti-mouse antibodies [HAMA] or heterophile antibodies), which may cause interference in some immunoassays. Caution should be used in interpretation of results and the laboratory should be alerted if the result does not correlate with the clinical presentation.
Liver-derived alkaline phosphatase (ALP) has some cross-reactivity in this assay: 100 U/L of liver ALP activity gives a result of 2.5 mcg/L to 5.8 mcg/L. Accordingly, serum specimens with significant elevations of liver ALP activity may yield elevated results.