This is a qualitative test reported as Negative, Positive, or Indeterminate based on the interpretation of four assay tube values. Numeric values of these tubes will also be reported.
• NIL: This value represents nonspecific reactivity produced by the patient specimen.
• MITOGEN minus NIL (MITOGEN-NIL): This value serves as the positive control for the patient specimen, demonstrating successful lymphocyte activity.
• TB1 minus NIL (TB1-NIL): This tube specifically detects CD4+ lymphocyte reactivity, specifically stimulated by the TB1 antigens.
• TB2 minus NIL (TB2-NIL): This tube detects both CD4+ and CD8+ lymphocyte reactivity stimulated by TB2 antigens.
Table: Guidance for the Interpretation of QGOLD Results
Nil (IU/mL)
| TB1 minus Nil (IU/mL)
| TB2 minus Nil (IU/mL)
| Mitogen minus Nil (mL)
| QuantiFERON-TB Gold Plus Result
| Interpretation
|
≤ 8.0
| ≥ 0.35 and ≥ 25% of the Nil
| Any
|
Any |
Positive
| M. tuberculosis infection
likely |
Any
| ≥ 0.35 and ≥ 25% of the Nil
|
< 0.35 OR ≥ 0.35 and <25% of Nil
| < 0.35 OR ≥ 0.35 and <25% of Nil
| ≥ 0.5
| Negative
| M. tuberculosis infection
NOT likely
|
< 0.35 OR ≥ 0.35 and <25% of Nil
| < 0.35 OR ≥ 0.35 and <25% of Nil
| < 0.5
|
Indeterminate | Likelihood of M. tuberculosis infection cannot be determined |
≥ 8.0
| Any
|
For more information regarding QuantiFERON-TB gold testing, including follow-up to positive results, see guidance from the Wisconsin Department of Health Services - Wisconsin Tuberculosis Program.